The standard “ISO 16900-2 Respiratory protective devices - Test methods and test equipment - Part 2: Determination of respiratory resistance” published by the International Organization for Standardization (ISO) describes respiratory resistance test methods for:

For the purposes of this standard, in addition to the terms and definitions given in “ISO 16972 Respiratory protective devices - Vocabulary and graphical symbols”, the following definitions apply:
To apply this standard, at least the following parameters must be specified in the performance standard:
Unless otherwise stated, values stated in this standard are expressed as nominal values. Values not stated as maximum or minimum, except for temperature limits, are subject to a tolerance of plus/minus 5 percent. Unless otherwise stated, ambient conditions for testing shall be between 16 degrees and 32 degrees and a relative humidity of 50 percent plus/minus 30 percent. Any temperature limits stated are subject to accuracy of plus/minus 1 degree.
For each of the required measurements performed in accordance with this standard, a corresponding estimate of the measurement uncertainty is evaluated. This uncertainty estimate is stated when reporting test results to enable the user of the test report to evaluate the reliability of the result in accordance with the annex to this standard (Annex A).
All flow velocity values shall be corrected to 1 013 hPa and 20 degrees. Measurement uncertainty is calculated in accordance with Joint Committee on Metrology Guidelines 100.
When performing “Method 1: Static inhalation resistance for filters and respiratory interfaces,” the filter or respiratory interface is mounted on a support and air is drawn through the device at a constant flow rate. The convention for reporting inhalation resistance is that if the pressure relative to the atmosphere within a respiratory interface or downstream of a filter is negative during the inhalation resistance test, the result is prefixed with a “-”. If the relative pressure within a respiratory interface is positive, the result is prefixed with a “+”. The pressure drop across a filter is reported without any prefix.
The main equipment used in this method includes:
The filter holder can be provided by the manufacturer of the respiratory protective device. It is important that the device support does not reduce the effective working area of the filter or respiratory protective device.
The procedure for respiratory protection device filters is as follows:
Our organization has a strong staff that closely follows the developments in the world in the field of science and technology and constantly improves itself. Among the numerous test, measurement, analysis and evaluation studies provided for businesses in various sectors, there are also test services in accordance with the standard “ISO 16900-2 Respiratory protective devices - Test methods and test equipment - Part 2: Determination of respiratory resistance”.