The “ISO 14801 Dentistry - Implants - Dynamic loading test for endosseous dental implants” standard, developed by the International Organization for Standardization (ISO) , describes a dynamic testing method for single-post endosseous dental implants of the transmucosal type, together with prefabricated prosthetic components.

This method is most useful for comparing endosseous dental implants of different designs or sizes. This international standard is not a test of the fundamental fatigue properties of the materials from which endosseous implants and prosthetic components are made. Furthermore, this standard is not applicable to dental implants or magnetic attachments with endosseous lengths shorter than 8 mm.
While this standard simulates the functional loading of an endosseous dental implant under worst-case conditions, it is not applicable for predicting the in vivo performance of an endosseous dental implant or denture, particularly if more than one endosseous dental implant is used for a denture.
For the purposes of this standard, the following terms and definitions apply:
In principle, testing should be performed on samples representative of the finished device (i.e. implant components that have undergone the same manufacturing process and sterilization as the device to be marketed). If the manufacturer intends the endosseous dental implant to be sterilized by the clinician prior to surgery, sterilization should be performed prior to testing as specified in the manufacturer's instructions for use. However, if there is evidence that the specified sterilization method does not have a significant effect on the properties of all materials in the tested samples, sterilization prior to testing is not required.
A multi-piece endosseous dental implant should be tested assembled according to its intended use. An endosseous dental implant component recommended by the manufacturer for use with components from another manufacturer should be tested assembled according to the manufacturer's statement. When a multi-piece device is assembled using screw connections, these should be used according to the manufacturer's recommendations and tightened to the manufacturer's recommended torque using the equipment provided with the implant system (implant screwdriver, torque wrench) or, if the original tools are not available, a device that provides torque within plus/minus 5 percent of the recommended value. The tightening sequence should be as recommended by the manufacturer.
If a component of an endosseous dental implant system is available in a variety of sizes or configurations, testing for worst-case conditions within the proposed use should be performed. The worst-case selection should be justified and documented. Guidance on how to select the worst-case is included in the standard appendix (Appendix B).
Among the numerous tests, measurements, analyses and evaluation studies that our organization has carried out for businesses in various sectors, it also provides testing services defined in the “ISO 14801 Dentistry - Implants - Dynamic loading testing for endosseous dental implants” standard with its trained and expert staff and advanced technological equipment.