The ISO 5356-1 Anesthesia and respiratory equipment - Conical connectors - Part 1: Cones and sockets standard, published by the International Organization for Standardization (ISO), describes the size and measurement requirements for cones and sockets designed to connect anesthetic and respiratory equipment (e.g., respiratory systems, anesthetic gas evacuation systems, and vaporizers). Therefore, cones and sockets are not considered devices in themselves.

This standard includes requirements for the following types of cones and sockets:
The ISO 5356-1 standard does not specify which medical devices and accessories these cones and sockets should be supplied with. Requirements regarding the application of cones and sockets are outside the scope of this standard, but are addressed in relevant international standards for specific medical devices and accessories. Requirements for screw-threaded weight-bearing connectors are found in the ISO 5356-2 standard.
General requirements are given in this standard as follows:
Compatibility is checked by fitting the appropriate plug or ring test gauge onto the cone or socket according to the shape and table, applying an axial force of 35 ± 3,5 N for 8,5 mm, 11,5 mm and 15 mm cones or sockets and 50 ± 5 N for 22 mm and 30 mm cones or sockets, while simultaneously rotating the cone or socket 20 degrees ± 5 degrees. The leading edge should be between the minimum and maximum diameter increments of the gauge. Cones, sockets and gauges should be kept at a temperature of 20 degrees ± 3 degrees during testing.
In the risky world of anesthesia and respiratory care, where a single incorrect connection can mean the difference between life and death, standardization is not just a convenience, but a lifeline. At the heart of this critical infrastructure is the ISO 5356-1 standard, which regulates conical connectors used in respiratory systems, anesthesia equipment, and ventilators. Regularly updated, this standard ensures the seamless and safe connection of medical devices from different manufacturers.
Connectors covered by the ISO 5356-1 standard are universal plugs and sockets for the respiratory world and provide the following:
This standard is compliant with regional regulations, including approval under MDR (EU) 2017/745 in Europe and FDA 21 CFR 868.5905 in the US.
Prior to the publication of this standard, hospitals faced a number of difficulties with incompatible equipment. A ventilator from Manufacturer A might not be compatible with a breathing circuit from Manufacturer B, which could lead to:
The ISO 5356-1 standard was developed for the following purposes:
Compliance for manufacturers and hospitals is ensured as follows:
This standard is critical because it prevents life-threatening incorrect connections in anesthesia and ventilators. This standard is considered important for the following reasons:
In short, the ISO 5356-1 standard is not just a standard, but also a life-saving global safety protocol for every operating room and intensive care unit.
In conclusion, the ISO 5356-1 standard is a silent safeguard in operating rooms and intensive care units worldwide. By mandating precision even in something as simple as a connector, it prevents chaos at the most critical moments of patient care. As mechanical ventilation evolves with AI-powered ventilators and home respiratory devices, the interoperability, safety, and reliability principles of this standard remain unquestionable.
For many years, our organization, with its competent and experienced staff, has been understanding the needs of its clients, providing the advanced testing services they require, and assisting them in the establishment, implementation, and improvement of management systems. Within this framework, we also provide testing services compliant with the ISO 5356-1 standard.