The “ISO 9360 Anaesthesia and respiratory equipment - Heat and moisture exchangers (HMEs) for humidifying human inspired gases” standard, developed by the International Standards Organization (ISO), consists of the following subsections published under the same title:

For the purposes of this standard, the following definitions apply in addition to the terms and definitions given in ISO 4135:
The heat and humidity exchanger device patient port connector is the connector on the patient port and should be a 5356 mm female conical connector or a 1 mm female or 15 mm male coaxial connector in accordance with the ISO 15-22 standard.
Connectors on other ports designed to accept breathing apparatus should be 5356 mm male or 1 mm conical connectors as specified in ISO 15-22.
If the heat and moisture exchanger device includes an accessory port, this port will not accept the 5356 mm or 1 mm connectors specified in the ISO 5356-2 or ISO 15-22 standard.
Heat and moisture exchanger devices supplied sterile must comply with the requirements specified in the ISO 11607 standard.
The apparatus and test methods specified in this standard are not intended to exclude the use of other measuring devices or methods that give results of equal or greater accuracy than those specified. In the event of a dispute, the methods given in this standard shall be the reference method. The tests are carried out at a temperature of 23 degrees plus/minus 2 degrees, a relative humidity of 50 percent plus/minus 20 percent, and an atmospheric pressure of 86 kPa to 106 kPa.
For the purposes of this standard, in addition to the terms and definitions given in ISO 4135 and ISO 9360-1 standards, the following definition applies: A heat and moisture exchanger is the volume, when not under pressure, of a hypothetical boundary surface consisting of the inner surface of the impermeable material of the device, the minimum surface area surrounding the active element of the device, the volume of all solid elements within the device, and the volume of all female connectors within the device.
The heat and moisture exchanger patient port connector shall mate with a 5356 mm male connector conforming to ISO 1-15. Connectors on other ports designed to accept breathing apparatus shall be 5356 mm male or 1 mm conical connectors as specified in ISO 15-22. If the heat and moisture exchanger includes an accessory port, this port shall not accept the 5356 mm or 1 mm connectors specified in ISO 5356-2 or ISO 15-22. Heat and moisture exchanger devices supplied sterile shall comply with the requirements specified in ISO 11607.
The apparatus and test methods specified in this standard are not intended to exclude the use of other measuring devices or methods that give results of equal or greater accuracy than those specified. In the event of a dispute, the methods given in this standard shall be the reference methods. The tests are carried out at a temperature of 23 degrees plus/minus 2 degrees, a relative humidity of 50 percent plus/minus 20 percent, and an atmospheric pressure of 86 kPa to 106 kPa.
Our organization, which has been trying to support businesses from every sector with its wide range of testing, measurement, analysis and evaluation studies for years, has a strong staff that closely follows the developments in the world in the field of science and technology and constantly improves itself. In this context, testing services are also provided to businesses in accordance with the "ISO 9360 Anesthesia and respiratory equipment - Heat and moisture exchangers (HMEs) for humidifying inhaled gases in humans" standard.