The standard "ASTM E1640-23 Standard test method for determining glass transition temperature by dynamic mechanical analysis" developed by the American Society for Testing and Materials (ASTM) describes a test method that involves determining the glass transition temperature (Tg) of materials using dynamic mechanical analyzers. This test method is applicable to thermoplastic polymers, thermoset polymers, and partially crystalline materials that are thermally stable in the glass transition region.

The temperature range for this test method depends on the instrumentation used, but the minimum temperature should be approximately minus 150 degrees to cover all materials.
This test method is designed for materials with elastic modulus in the range of 0,5 MPa to 100 GPa.
Specific technical terms used in this standard are included in “ASTM D4092-21 Standard terminology for plastics: dynamic mechanical properties” and “ASTM E1142-15 Standard terminology related to thermophysical properties”, including Celsius, dynamic mechanical analyzer, glass transition, glass transition temperature, loss modulus, storage modulus, tangent delta, and viscoelasticity.
In conducting tests, a sample of known geometry is placed in mechanical oscillation at a constant or resonant frequency and changes in the viscoelastic response of the material are monitored as a function of temperature. Under ideal conditions, during heating, the glass transition region is marked by a rapid decrease in the storage modulus and a rapid increase in the loss modulus and tangent delta.
The glass transition of the test sample is indicated by the extrapolated onset of decrease in storage modulus, which marks the transition from a glassy solid to a rubbery solid.
This test method can be used to find the glass transition region of amorphous and semicrystalline materials and assign the glass transition temperature. Dynamic mechanical analyzers provide a way to measure these changes by monitoring changes in the viscoelastic properties of a material as a function of temperature and frequency. In ideal cases, the onset temperature of the decrease in storage modulus marks the glass transition.
The glass transition occurs over a temperature range. This method assigns a single temperature (Tg) to represent the temperature range measured by dynamic mechanical analysis. Tg can be determined by a variety of techniques and can vary with that technique.
A glass transition temperature (Tg) is useful in characterizing many important physical properties of thermoplastic, thermoset, and semicrystalline materials, including their thermal history, processing conditions, physical stability, progression of chemical reactions, degree of cure, and both mechanical and electrical behavior.
This test method is found useful for quality control, specification acceptance and research.
Since sample size will usually be small, it is essential that each sample be homogeneous or representative of the material as a whole, or both. An increase or decrease in heating rates from those specified may alter the results.
A transition temperature is a function of the experimental frequency, so the test frequency should always be specified. (Transition temperature increases with increasing frequency). Extrapolation to a common frequency can be accomplished by using a predetermined frequency shift factor or by assuming a frequency shift factor of approximately 8 degrees for each decade of frequency. Such extrapolations should be reported.
The function of the apparatus used in these tests is to hold a sample of uniform size in such a way that the sample acts as an elastic and dispersive element in a mechanically oscillating system. Dynamic mechanical analyzers generally operate in one of several modes.
This device includes:
Among the numerous testing, measurement, analysis and evaluation studies that our organization has carried out for businesses in various sectors, it also provides testing services defined in the "ASTM E1640-23 Standard test method for determining glass transition temperature by dynamic mechanical analysis" standard with its trained and expert staff and advanced technological equipment.